dc.contributor.author |
Yıldırım, Ender
|
|
dc.contributor.author |
Arıkan, M. A. Sahir
|
|
dc.contributor.author |
Kulah, Haluk
|
|
dc.date.accessioned |
2020-05-12T05:05:36Z |
|
dc.date.available |
2020-05-12T05:05:36Z |
|
dc.date.issued |
2011 |
|
dc.identifier.citation |
Yildirim, E.; Kulah, H.; Arikan, M.A.S.,"An Electrostatically Actuated Parylene Microvalve for Lab-On-A-Chip Applications", 2011 16th International Solid-State Sensors, Actuators and Microsystems Conference, Transducers'11, pp. 250-253, (2011). |
tr_TR |
dc.identifier.isbn |
978-145770157-3 |
|
dc.identifier.uri |
http://hdl.handle.net/20.500.12416/3720 |
|
dc.description.abstract |
This paper presents a novel electrostatic microvalve
to control in-plane flow on parylene based
lab-on-a-chip-devices. Normally-closed design of the
microvalve insulates the working fluid from the electric
field, while providing low leakage up to 40 kPa inlet
pressure. Prototypes are fabricated and tested for pull-in
and flow characterization. Pull-in voltage is measured to
be 150 V independent of the working fluid. No leakage is
detected up to 20 kPa inlet pressure. |
tr_TR |
dc.language.iso |
eng |
tr_TR |
dc.relation.isversionof |
10.1109/TRANSDUCERS.2011.5969441 |
tr_TR |
dc.rights |
info:eu-repo/semantics/openAccess |
tr_TR |
dc.subject |
Microvalve |
tr_TR |
dc.subject |
Parylene, |
tr_TR |
dc.subject |
Electrostatic, |
tr_TR |
dc.subject |
In-plane |
tr_TR |
dc.title |
An Electrostatically Actuated Parylene Microvalve for Lab-On-A-Chip Applications |
tr_TR |
dc.type |
conferenceObject |
tr_TR |
dc.relation.journal |
2011 16th International Solid-State Sensors, Actuators and Microsystems Conference, Transducers'11 |
tr_TR |
dc.contributor.authorID |
31835 |
tr_TR |
dc.identifier.startpage |
250 |
tr_TR |
dc.identifier.endpage |
253 |
tr_TR |
dc.contributor.department |
Çankaya Üniversitesi, Mühendislik Fakültesi, Makine Mühendisliği Bölümü |
tr_TR |